Pamas S50P Fuel

+courtesy of Pamas 

PAMAS S50P FuelThe online particle counter PAMAS S50P Fuel is used asa online measuring instrument for condition monitoring and contamination analysis of liquid fuel(e.g. diesel). The instrument measures the number of particles in eight different size channels and can function as early alerts in case of failures caused by bigger particle sizes (e.g. corrosion). This makes PAMAS S50P Fuel indispensable for preventive condition monitoring of liquid fuel, as it prevents the filling of contaminated fuel into storage tanks. On its 6-digit LED display, the instrument reports the triple code according to ISO 4406. The PAMAS S50P Fuel is equipped with a pump to control the flow rate to 25 ml/min at a pressure range from 0 to 6 bar.

 

 

 

 

Application                                                          Standards and Methods

Contamination control and condition monitoring of liquid fuel          ISO11171  ISO 4406 SAE4059

Sample input

Pressure range: 0 to 6 bar
Sample flow rate: 25 ml/min

Volumetric particle sensor PAMAS HX

Calibration range: 4 to 70 µm(c) as per calibration standard ISO 11171
Measurement in eight size channels: > 4 µm(c), > 6 µm(c), > 10 µm(c), > 14 µm(c), > 21 µm(c), > 25 µm(c), > 38 µm(c) and > 70 µm(c)
Maximum particle concentration: 24.000 particles per ml at a coincidence quote of 7.8% and a flow rate of 25 ml/min

Standardised measuring results

Report of measuring results according to cleanliness class standards SAE AS 4059 and ISO 4406.
The particle counter measures ISO classes from 0 to 22.

Interfaces for data transfer

Standard equipment: RS485
Optional equipment: analogue 4-20 mA interface. Parallel data transmission for the size channels > 4, > 6, > 14 and > 70 µm(c) or serial data transmission for all eight size channels.

Software

PAMAS POV (PAMAS Online Visualisation): Software for online visualisation of measuring results and for long-term trend monitoring